RegWardenFDA facility compliance records
Recall

Class II — GE Medical Systems Information Technologies Inc

Recalling firmGE Medical Systems Information Technologies Inc
ClassificationClass II
StatusOngoing
ReasonSoftware issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
ProductPORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209
Quantity23 units
DistributionUS: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
Initiated2026-07-20
Classified2026-08-27
Recall numberZ-3036-2026
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — GE Medical Systems Information Technologies Inc (Z-3036-2026).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/99536/

Not open yet

Alerts for GE Medical Systems Information Technologies Inc

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.