Xoran Technologies LLC
6 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (6)
Registered facilities (1)
- Xoran Technologies, LLC FEI 3004198450 · MI
Recalls & enforcement
- Class IITerminated2014-09-08Out of tolerance kVp, Unexposed (black) frames, and Out of tolerance Dose.
- Class IITerminated2010-07-27A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints receive…
- Class IITerminated2010-07-27The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the…
- Class IITerminated2010-07-27The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.
- Class IITerminated2010-07-27A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints receive…
- Class IITerminated2009-12-14Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “Xoran Technologies LLC — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-10028111-xoran-technologies-llc/
Alerts for Xoran Technologies LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.