Class II — Xoran Technologies, Inc.

Recalling firmXoran Technologies LLC
Registered facilityXoran Technologies, LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha
ProductXoran Minicat00158
QuantityN/A
DistributionWorldwide Distribution - US nationwide
Initiated2010-07-27
Classified2013-10-25
Terminated2014-09-11
Recall numberZ-0039-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Xoran Technologies, Inc. (Z-0039-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66338/

Not open yet

Alerts for Xoran Technologies LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.