RAYE'S, Inc dba Sizewise Manufacturing a wholly owned subsid
1 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (1)
Registered facilities (5)
- Agiliti Health - Ellis FEI 3002693132 · KS
- Agiliti Health - Corona FEI 3000204953 · CA
- Agiliti Health - Milwaukee FEI 3029836575 · WI
- Agiliti Health - Hays FEI 1931307 · KS
- RASMUSSEN DE TECATE S.A. DE C.V. FEI 3034243939 · MX
Recalls & enforcement
- Class IIOngoing2024-01-25Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a…
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.