RegWardenFDA facility compliance records
Recall

Class II — Raye's Inc.

Recalling firmRAYE'S, Inc dba Sizewise Manufacturing a wholly owned subsid
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonRetrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
ProductSW Bari Lift & Transfer, Model Number 38060000
Quantity1,149 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, UK.
Initiated2024-01-25
Classified2024-04-18
Recall numberZ-1595-2024
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Raye's Inc. (Z-1595-2024).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/93887/

Not open yet

Alerts for RAYE'S, Inc dba Sizewise Manufacturing a wholly owned subsid

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.