ZEUS SCIENTIFIC LLC
3 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (3)
Registered facilities (1)
- ZEUS SCIENTIFIC LLC FEI 2242436 · NJ
Recalls & enforcement
- Class IIOngoing2023-10-31The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
- Class IITerminated2019-08-12FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.
- Class IITerminated2019-07-22Assay files used on the open Dynex DSX Instrument to process IVD ELISA samples-programming error results in well H6 being skipped for the addition of the TMB substrate. Well H6 will generate a negative result regardless of whether the patient was negative or positive for that ant…
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “ZEUS SCIENTIFIC LLC — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-2242436-zeus-scientific-llc/
Alerts for ZEUS SCIENTIFIC LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.