RegWardenFDA facility compliance records
Facility

ZEUS SCIENTIFIC LLC

3 recall(s) on record — most severe on record: Class II.

Activity by year

FDA actions on record by year2016: 1 inspections; 2019: 2 recalls; 2020: 1 inspections; 2023: 1 recalls120162017201822019120202021202212023202420252026
  • Recalls (3)
  • Inspections (2)

Devices registered here (1)

  • Enzyme Linked Immunosorbent Assay, Herpes Simplex Virus, Hsv-1
    Product code MXJ (opens in a new tab) · Class II · Microbiology · 21 CFR 866.3305

Recalls & enforcement

  • Class IIOngoing2023-10-31
    The organization completed a product improvement process to the conjugate found within the certain ELISA kits. This modification was not reviewed and approved by FDA prior to its implementation.
  • Class IITerminated2019-08-12
    FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.
  • Class IITerminated2019-07-22
    Assay files used on the open Dynex DSX Instrument to process IVD ELISA samples-programming error results in well H6 being skipped for the addition of the TMB substrate. Well H6 will generate a negative result regardless of whether the patient was negative or positive for that ant…

FDA inspections (2)

  • NAI2020-08-31
    No Action Indicated · Postmarket Assurance: Devices; Project Evaluation: Devices
  • VAI2016-02-19
    Voluntary Action Indicated · Bioresearch Monitoring · citations posted

Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “ZEUS SCIENTIFIC LLC (FEI 2242436) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/2242436-zeus-scientific-llc/

Not open yet

Alerts for ZEUS SCIENTIFIC LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.