ALPHATEC HOLDINGS, INC.

17 recall(s) on record — most severe on record: Class II.

Recalls by year

FDA actions on record by year2010: 1 recalls; 2011: 1 recalls; 2012: 2 recalls; 2013: 2 recalls; 2014: 1 recalls; 2015: 1 recalls; 2016: 1 recalls; 2017: 1 recalls; 2021: 1 recalls; 2022: 1 recalls; 2023: 1 recalls; 2024: 2 recalls; 2026: 2 recalls1’101’112’122’131’141’151’161’17’18’19’201’211’221’232’24’252’26
  • Recalls (17)

Registered facilities (3)

Recalls & enforcement

20262 records
  • Class IIOngoing2026-05-13
    Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring sy…
  • Class IIOngoing2026-01-02
    Due a design issue where the navigated array connection geometry is incorrect.
20242 records
20231 record
  • Class IIOngoing2023-07-07
    Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to ove…
20221 record
  • Class IIOngoing2022-09-14
    Due to manufacturing instruments which included additional features and adequate consideration was not given to the need for a new Regulatory filing.
20211 record
  • Class IITerminated2021-04-26
    There is a potential for the screw blocking mechanism on the anterior cervical plate system to disassociate intraoperatively/postoperatively or unlock postoperatively.
20171 record
  • Class IITerminated2017-05-05
    Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.
20161 record
  • Class IITerminated2016-10-05
    Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.
20151 record
  • Class IITerminated2015-04-10
    During a field audit it was discovered by an employee that a 16mm Trestle Luxe Screw has been both laser marked and labeled as a 14mm Trestle Luxe Screw (4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm). As a result the firm recalling the affected lots codes for this mi…
20141 record
  • Class IITerminated2014-06-26
    Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use.
20132 records
  • Class IITerminated2013-06-28
    The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide…
  • Class IITerminated2013-05-06
    Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head …
20122 records
  • Class IITerminated2012-09-25
    The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.
  • Class IITerminated2012-05-29
    On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
20111 record
  • Class IITerminated2011-07-06
    The rods were incorrectly laser etched as ILLICO® Straight CP Tl Rods, 110mm, when they were in fact manufactured to 100mm in length.
20101 record
  • Class IITerminated2010-12-09
    The recall was initiated because Alphatec Spine confirmed that the TI Standard Polyaxial Screw Assembly was laser marked with the incorrect length of 45 MM.

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.

Cite this record

RegWarden. “ALPHATEC HOLDINGS, INC. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-9072657-alphatec-holdings-inc/

Not open yet

Alerts for ALPHATEC HOLDINGS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.