Class II — Alphatec Spine, Inc.
| Recalling firm | ALPHATEC HOLDINGS, INC. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional. |
| Product | ILLlCO® MIS Posterior Fixation System, Bone-screw Internal Spinal Fixation System, Orthosis, Spinal Pedical Fixation Orthosis, Spondylolisthesis Spinal Fixation The ILLICO MIS Posterior Fixation System is intended to facilitate the surgical correction of noncervical spinal deformities by providing temporary internal fixation and stabilization during bone graft healing and/or fusion mass development. When used for a minimally invasive posterior approach ILLlCO® MIS Instrumentation is used in conjunction with polyaxial screw components. The ILLICO MIS Posterior Fixation System is intende… |
| Quantity | 209 screws |
| Distribution | Nationwide Distribution including AZ, IL, CA, SC, TN, TX, GA, CT, NY, FL, OK, LA, MO, NV,and UT. |
| Initiated | 2012-09-25 |
| Classified | 2012-10-29 |
| Terminated | 2013-03-29 |
| Recall number | Z-0168-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Alphatec Spine, Inc. (Z-0168-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63385/