RegWardenFDA facility compliance records
Facility

Hologic, Inc.

41 recall(s) on record — most severe on record: Class I.

Activity by year

FDA actions on record by year2012: 1 recalls; 2013: 2 recalls; 2014: 2 recalls; 2016: 5 recalls, 2 inspections; 2017: 4 recalls; 2018: 2 recalls, 1 inspections; 2019: 1 recalls; 2020: 2 recalls; 2021: 1 recalls; 2022: 4 recalls, 1 inspections; 2023: 4 recalls; 2024: 6 recalls, 1 inspections; 2025: 4 recalls; 2026: 3 recalls1’122’132’14’157’164’173’181’192’201’215’224’237’244’253’26
  • Recalls (41)
  • Inspections (5)

Devices registered here (2)

  • Parainfluenza Multiplex Nucleic Acid Assay
    Product code OOU (opens in a new tab) · Class II · Microbiology · 21 CFR 866.3980
  • Real Time Nucleic Acid Amplification System
    Product code OOI (opens in a new tab) · Class II · Clinical Chemistry · 21 CFR 862.2570

Recalls & enforcement

20263 records
  • Class IIOngoing2026-04-08
    Complaints have been received of systems developing loose, missing, or broken internal bolts over time.
  • Class IIOngoing2026-03-23
    It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibratio…
  • Class IIOngoing2026-02-18
    Due to product exhibiting potential to generate either invalid or false negative results.
20254 records
  • Class IIOngoing2025-12-15
    Presence of particulates in affected devices that can be deposited into breast tissue during use.
  • Class IIOngoing2025-12-12
    Fungal contamination of affected lot with Parengyodontium album.
  • Class IIOngoing2025-01-15
    Devices presented a condition in which the inner needle separated from the hub and produced a longer throw of the inner needle or dislodgement.
  • Class IIOngoing2025-01-08
    Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN…
20246 records
  • Class IIOngoing2024-12-13
    Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.
  • Class IIOngoing2024-11-08
    SARS-CoV-2/Flu A/B/RSV Assay Kit may erroneously produce SARS-CoV-2 false positive results.
  • Class IOngoing2024-10-10
    Firm received reports of adverse events/complications that include serious injuries, such as pain, infection, rash, device migration, device erosion, seroma, and discomfort.
  • Class IIOngoing2024-08-19
    Identified an issue with Selenia Dimensions 1.12.0 and 3Dimensions 2.3.0 system software when used in combination with C-View software in which the quality of the final C-View synthesized 2D image is impacted and may appear blurry and result in misdiagnosis with the possibility b…
  • Class IIOngoing2024-06-13
    Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA System exceeded the limit for acc…
  • Class IOngoing2024-03-13
    Safety Information: Complaints describe complications/adverse events that include pain, infection, rash, device migration, device erosion, seroma, discomfort, or other complications from feeling the device in the breast, and the need for additional medical treatment to remove the…
20234 records
20224 records
  • Class IICompleted2022-10-10
    IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
  • Class IIOngoing2022-08-09
    Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on t…
  • Class IIOngoing2022-03-18
    Samples, run on a GBS Assay, that have analyte concentrations that are at or near the limit of detection may produce false negative results.
  • Class IIOngoing2022-01-27
    Failed stability specifications for reagent kits that may result in false negative test results.
20211 record
  • Class IIOngoing2021-02-26
    There is a potential for tube trays to leak which could potentially result in false positive results, false negative results or invalid patient results.
20202 records
  • Class IITerminated2020-08-04
    Tube trays may leak which could invalidate patient results or assay worklists, causing delayed results for any assays with which they are used. A possibility exists of false negative result with flu A/B/RSV assay or paraflu assay.
  • Class IITerminated2020-01-21
    The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.
20191 record
  • Class IIOngoing2019-10-10
    Reagent component included in flu test kit may cause both high background signal in a true negative sample leading to false positive and/or false negative results. This could result in false positive and/or false negative results on patient samples or failed controls leading to i…
20182 records
  • Class IITerminated2018-12-06
    Calibration issue not possible to visualize contrast uptake in the subtracted mammography images
  • Class IITerminated2018-01-20
    The Panther or the Panther Fusion glass panel separates from the 19 inch touch screen monitor and could potentially fall and hit an operator.
20174 records
  • Class IITerminated2017-11-29
    Rapid fFN cassettes were manufactured with material (A137 ascites) from a new supplier (SDIX). The products were distributed to customers before FDA approved the change to use the material from SDIX.
  • Class IITerminated2017-09-15
    An incorrect incubator was installed on one Panther System which potentially led to incorrect Aptima Combo 2(R) assay (AC2) results.
  • Class IITerminated2017-07-07
    The firm is implementing a product correction involving its Affirm® Lateral Arm Upright Biopsy Accessory, ASY-09880. The correction has been initiated due to a component (the Left-hand Guide or the Blue Needle Guide) issue that could impact needle alignment for all the left-appro…
  • Class IITerminated2017-06-27
    A Tomcat Instrument software anomaly that causes the pipettor arm to make physical contact with the mucous shelf which could potentially lead to incorrect patient results.
20165 records
20142 records
  • Class IITerminated2014-05-13
    Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
  • Class IITerminated2014-01-24
    NovaSure Radiofrequency Control Units may not meet a requirement of the electrical standard, IEC 60601-1,
20132 records
  • Class IITerminated2013-09-06
    Software: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most recently dated images
  • Class IITerminated2013-04-25
    The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.
20121 record
  • Class IITerminated2012-09-14
    Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180® around vertical axis as well as 180® around a horizontal axis…

FDA inspections (5)

  • NAI2024-08-13
    No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices
  • OAI2022-10-14
    Official Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted
  • NAI2018-12-10
    No Action Indicated · Project Evaluation: Devices
  • NAI2016-12-07
    No Action Indicated · Bioresearch Monitoring
  • NAI2016-06-30
    No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices

Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Hologic, Inc. (FEI 2024800) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/2024800-hologic-inc/