EOS imaging
7 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (7)
Devices registered here (1)
- System, Image Processing, RadiologicalProduct code LLZ (opens in a new tab) · Class II · Radiology · 21 CFR 892.2050
Recalls & enforcement
- Class IITerminated2020-12-22Inadequate images resizing and 2D measurement errors may occur when biplanar acquisition has been performed with patient orientation different from AP (Antero-Posterior).
- Class IITerminated2018-08-013D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two images, where one of the images has been manually flipped. Projection errors may lead to errors in calculation of 3D clinical parameters when ad…
- Class IITerminated2017-05-30ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the operator must examine the vacuum safety indicator to check the vacuum is monitored and sufficient and the grab handle correctly attached. Do…
- Class IITerminated2015-02-17When performing calibration, an alert message on the spectral filtration of the X-ray beam may be suppressed. Improper filtration of the X-ray Beam can then occur in exams set up with copper filtration.
- Class IITerminated2013-11-29Error copying information to the clinical file and to the report when two 3D models are opened simultaneously on the sterEOS workstation
- Class IITerminated2013-11-07EOS imaging discovered during production internal testing that the X-ray beam collimation set up by the operator at the user interface of the acquisition workstation is different from the actual collimation of the X-ray beam operated by the EOS system during the X-ray acquisitio…
- Class IITerminated2012-06-19It was discovered of several generator failures of the EOS System due to failure of its anode controller board that drives the rotation of the X-ray tube anode.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “EOS imaging (FEI 3006165260) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3006165260-eos-imaging/
Alerts for EOS imaging
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.