Recall

Class II — Eos Imaging Inc

Recalling firmEOS imaging
Registered facilityEOS imaging (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonWhen performing calibration, an alert message on the spectral filtration of the X-ray beam may be suppressed. Improper filtration of the X-ray Beam can then occur in exams set up with copper filtration.
ProductEOS, Digital radiography system used in general radiographic examinations.
Quantity13 units installed in US
DistributionUS Distribution to the states of: CA, DE, PA, MN, FL, MO, OH, IN and IL.
Initiated2015-02-17
Classified2015-05-05
Terminated2017-05-31
Recall numberZ-1460-2015
InitiationVoluntary: Firm initiated