RMO
7 recall(s) on record — most severe on record: Class I.
Activity by year
- Recalls (7)
Devices registered here (1)
- Ventilator, Non-Continuous (Respirator)Product code BZD (opens in a new tab) · Class II · Anesthesiology · 21 CFR 868.5905
Recalls & enforcement
- Class IOngoing2026-06-25Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
- Class IOngoing2023-11-20Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about …
- Class IOngoing2023-09-13If ventilator is on internal battery, not intended to serve as a primary power source, low/critically low battery alarms will sound, but a fault leads to sudden power loss. If power fails, then ventilation stops and a Total Power Failure (TPF) alarm should sound, but it's powered…
- Class ICompleted2019-12-12Combination of software and a component failure may cause audible alarms not to operate properly, the alarm buzzer not work, for ventilators that have a failed electronic component and, that are stored without AC power connected for more than 36 hours leading to full depletion of…
- Class IITerminated2016-07-12Masks in this lot, labelled and sold as 62702 vented Quattro Air FFM size Medium actually contain a non-vented (NV) Quattro Air FFM Large mask.
- Class IITerminated2016-01-13Resmed Corporation is recalling Astral 100, 150 External Battery because of a start-up issue.
- Class IITerminated2015-05-05The Astral device allows clinicians to disable all alarms including those that detect circuit disconnection. Inappropriate disabling of alarms in ventilator dependent patients can lead to hazards of insufficient ventilation and insufficient expiratory pressure which can result i…
Related
- All facilities operated by RESMED CORP.
- RESMED PTY LTD. (AU)
- RESMED CORP (CA)
- ResMed Asia Pte. Ltd. (SG)
- ResMed Malaysia Operations Sdn Bhd (MY)
- ResMed Corp (GA)
- ResMed West Coast Warehouse (CA)
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “RMO (FEI 3030970580) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3030970580-rmo/
Alerts for RMO
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.