RegWardenFDA facility compliance records
Recall

Class II — Hamilton Medical, Inc.

Recalling firmHAMILTON MEDICAL, INC.
Registered facilityHAMILTON MEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonAn issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.
ProductHamilton Medical Infant Flow Sensor, single use, (3.1m), Part number 260179. The infant flow sensor is an accessory for the Hamilton Medical Ventilators. Product Usage: The infant flow sensor is a device to measure patient air flow and pressures within the Hamilton Medical Ventilators.
Quantity154 total, all part numbers
DistributionUS Nationwide Distribution.
Initiated2015-06-01
Classified2015-07-02
Terminated2015-08-12
Recall numberZ-1965-2015
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Hamilton Medical, Inc. (Z-1965-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/71434/

Not open yet

Alerts for HAMILTON MEDICAL, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.