HAMILTON MEDICAL, INC.
10 recall(s) on record — most severe on record: Class I.
Recalls by year
- Recalls (10)
Registered facilities (1)
- HAMILTON MEDICAL, INC. FEI 3016723884 · NV
Recalls & enforcement
- Class IOngoing2023-07-26Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
- Class IOngoing2023-06-15Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
- Class IIOngoing2017-06-20Issue related to the performance of the motor in recalled product. During use, motor may cease to function.
- Class IITerminated2016-09-26Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.
- Class IITerminated2016-03-09After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.
- Class IITerminated2015-06-01An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation d…
- Class IITerminated2015-03-24Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
- Class ITerminated2014-04-22Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).
- Class IITerminated2013-09-26Manufacturer defined limits for improperly functioning loudspeakers of the HAMILTON C2 alarm system were exceeded.
- Class ITerminated2012-12-19Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin t…
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “HAMILTON MEDICAL, INC. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-2937708-hamilton-medical-inc/
Alerts for HAMILTON MEDICAL, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.