Class I — Hamilton Medical, Inc.
| Recalling firm | HAMILTON MEDICAL, INC. |
|---|---|
| Registered facility | HAMILTON MEDICAL, INC. (the firm's only registered facility) |
| Classification | Class I |
| Status | Ongoing |
| Reason | Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation. |
| Product | Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator |
| Quantity | 1216 units |
| Distribution | Distribution US nationwide, including Puerto Rico. OUS: Not provided |
| Initiated | 2023-07-26 |
| Classified | 2023-09-20 |
| Recall number | Z-2527-2023 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Hamilton Medical, Inc. (Z-2527-2023).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/92873/
Alerts for HAMILTON MEDICAL, INC.